PT - JOURNAL ARTICLE AU - Rogers, Patrick Scott AU - Pickett, Gwynedd AU - Vandorpe, Robert AU - Volders, David TI - Coil migration through two flow-diverting stents AID - 10.1136/bcr-2023-256863 DP - 2024 Jan 01 TA - BMJ Case Reports PG - e256863 VI - 17 IP - 1 4099 - http://casereports.bmj.com/content/17/1/e256863.short 4100 - http://casereports.bmj.com/content/17/1/e256863.full SO - BMJ Case Reports2024 Jan 01; 17 AB - We report a case of a patient who initially presented with a subarachnoid haemorrhage secondary to a ruptured supraclinoid internal carotid artery (ICA) blister aneurysm. The patient was treated successfully with a flow diverter stent (FD) and coiling; however, a large aneurysm recurrence via a feeding posterior communicating artery (PCOM) was noted on the 1-year follow-up angiogram. During the retreatment, a second FD in the ICA resulted in insufficient aneurysm stasis. Therefore, the decision was made to coil sacrifice the PCOM via posterior circulation access. During the first coil deployment, the distal coil end migrated through the mesh of two overlapping FD into the middle cerebral artery. This complication was a previously unrecognised possibility given the composition of the FD. This case report aims to discuss this process as a potential complication during neurointerventional procedures using these devices.