- Published on: 2 October 2018Response to case report; cephalad migration of externally powered spinal cord stimulator.Show More
Sirs, As a neuromodulator and instructor with extensive (30 year) experience with both "wired " and "wireless " systems, I feel compelled to repond to the conclusions of the cited case report.
While the authors mention lead migration as an known complication of spinal cord stimulation, it is useful to consider that Cameron (1) cited 361 lead migration events in 2753 patients , a 13.2% incidence. Anderson (2) , i...