A simpler approach to seal severe coronary perforation with bare metal stent

BMJ Case Rep. 2010 Oct 1:2010:bcr0620091954. doi: 10.1136/bcr.06.2009.1954.

Abstract

The case of a coronary artery perforation in which overinflation of a balloon at an angulated segment of the left anterior descending (LAD) artery after stent deployment resulted in an Ellis type III coronary artery perforation is presented. A bare metal stent (BMS) was used successfully to seal this high-grade perforation. Here, it is demonstrated that it may not be illogical to consider BMS as the first choice before embarking on use of a covered stent if the clinical/haemodynamic condition of the patient allows it and if a covered stent is not available.

Publication types

  • Case Reports

MeSH terms

  • Aged
  • Angioplasty, Balloon, Coronary / adverse effects*
  • Angioplasty, Balloon, Coronary / instrumentation*
  • Coronary Angiography
  • Coronary Vessels / injuries*
  • Echocardiography
  • Follow-Up Studies
  • Humans
  • Male
  • Myocardial Infarction / diagnostic imaging
  • Myocardial Infarction / therapy*
  • Stents / adverse effects*
  • Thrombolytic Therapy