Suspicious breast pseudocalcifications
Free
  1. Nuno André Almeida Costa1,
  2. Gisela Rio2,
  3. João Garrido Santos1
  1. 1Radiology, Instituto Portugues de Oncologia do Porto Francisco Gentil, EPE, Porto, Portugal
  2. 2Radiology, Hospital de Braga, Braga, Portugal
  1. Correspondence to Dr Nuno André Almeida Costa, nunoada{at}hotmail.com

Statistics from Altmetric.com

Request Permissions

If you wish to reuse any or all of this article please use the link below which will take you to the Copyright Clearance Center’s RightsLink service. You will be able to get a quick price and instant permission to reuse the content in many different ways.

Description 

We present a follow-up case of an invasive right ductal breast carcinoma in a 72-year-old woman.

A follow-up mammography study (figure 1) performed 1 year after right lumpectomy (wide local excision) shows postsurgical changes in the right breast and reveals ‘de novo’ coarse/amorphous microcalcifications in the left breast axis. Some of these calcifications show radiolucent centre, and can only be depicted in the craniocaudal mammographic images, raising concern for hypothetic cutaneous location, although presenting suspicious features.

Figure 1

Mammographic images showing postlumpectomy (wide local excision) changes in the right upper outer quadrant and coarse and amorphous calcifications in the left breast. Blue square shows the magnification of the calcifications, revealing some calcifications with radiolucent centre.

Cancer biomarkers and clinical status were unremarkable.

An ultrasound (US) evaluation was further advised. The patient was told not to apply any cream, lotion or deodorant on the day of the US study. Two weeks later, the patient was referred for the US examination. The evaluation was unremarkable. On the same day, a mammography study was repeated (figure 2) revealing total absence of the calcifications previously seen.

Figure 2

Mammographic images obtained 2 weeks after revealing the total absence of the calcifications—pseudocalcifications due to previous application of zinc-based ointment on the breast sulcus.

The patient reveals previous use of a zinc-based ointment on the breast sulcus, which she applied on the day of the first mammographic study. Ointment, deodorant and talcum may contain radiopaque components such as zinc, aluminium and magnesium, which may simulate calcifications at mammography.1 2

Artefacts reduce the quality of mammograms and may mimic or obscure abnormalities and cause interpretation errors.3

This case highlights the importance of obtaining high-quality mammographic images for accurate diagnosis as mammographic artefacts may obscure the true abnormalities or create pseudolesions.

Learning points

  • Ointment, deodorant and talcum may contain radiopaque components such as zinc, aluminium and magnesium, which may simulate calcifications at mammography.

  • To prevent these artefacts, technologists should inform the patients, in advance, not to wear powder, deodorant and body cream or ointment prior to the mammographic examination.

  • Mammographic artefacts may obscure the true abnormalities or create pseudolesions.

References

View Abstract

Footnotes

  • Contributors NAAC, GR and JGS: substantial contributions to the conception or design of the work, or the acquisition, analysis or interpretation of data. NAAC: drafting the work or revising it critically for important intellectual content; final approval of the version published; agreement to be accountable for all aspects of the work in ensuring that questions related to the accuracy or integrity of any part of the work are appropriately investigated and resolved.

  • Funding The authors have not declared a specific grant for this research from any funding agency in the public, commercial or not-for-profit sectors.

  • Competing interests None declared.

  • Patient consent Obtained.

  • Provenance and peer review Not commissioned; externally peer reviewed.