Article Text
Summary
Inferior vena caval (IVC) filters are used as a mechanical option for the treatment of venous thromboembolism (VTE) when standard anticoagulation therapy is either contraindicated or VTE recurs despite adequate anticoagulation. Filters are not without risk, however. Reported complications include filter migration and fracture of filter components, leading to IVC rupture and penetration into pericaval tissues (notably kidney, heart, pericardium, thoracic cavity, liver, bowel and aorta). Here we describe an extreme case of multiple organ penetration by a standard Celect caval filter.
- Vascular surgery
- Interventional radiology
Statistics from Altmetric.com
Footnotes
Contributors MMK: Data collection, drafting the article and final approval of the version to be published. NT and AS: Critical revision of the article. AH: Conception and critical revision of the article.