Article Text
Summary
This case report describes a successful hysteroscopic sterilisation using the Essure Permanent Birth Control device (Conceptus Inc., Mountain View, California, United States) after a failed procedure of the Adiana Permanent Contraception system (Hologic, Inc., Bedford, Maryland, United States). The delivery catheter of the Adiana system was able to be inserted into the left fallopian tube without difficulty and per manufacturer specifications. However, the position detection array was unable to sense four-quadrant tissue contact. The same issue occurred at the contralateral tube. Using the Essure system, the coils were able to be placed in both ostia easily and adequately. In patients in whom the Adiana system fails to occlude the fallopian tubes due to procedural, anatomic or device-related factors, the Essure procedure may be an efficient alternative.
Statistics from Altmetric.com
Footnotes
-
Competing interests None.
-
Patient consent Obtained.