Distal pacemaker lead fracture: a rare entity
Free
  1. D X Augustine1,
  2. K Carson1,
  3. A Garg2
  1. 1Department of Cardiology, RUH Bath NHS Foundation Trust, Bath, UK
  2. 2RUH Bath NHS Foundation Trust, Bath, UK
  1. Correspondence to D X Augustine, dxaugustine{at}hotmail.com

Statistics from Altmetric.com

Request Permissions

If you wish to reuse any or all of this article please use the link below which will take you to the Copyright Clearance Center’s RightsLink service. You will be able to get a quick price and instant permission to reuse the content in many different ways.

Description

An 87-year-old woman from a nursing home with a history of ventricular rate responsive demand (VVIR) Microny II SR (St. Jude Medical AB, Järfälla, Sweden) pacemaker insertion for syncope with complete heart block in 2003 was admitted following an episode of loss of consciousness. This occurred as she was being helped to the toilet by two carers each supporting her under one arm.

On admission her pulse was 40 beats/min. Her admission electrocardiogram (ECG) (figure 1) demonstrated intermittent ventricular pacing spikes failing to capture with an underlying rhythm of complete heart block and a ventricular rate of 40 beats/min. Her chest radiograph (figure 2, left panel) showed a distal ventricular pacing lead fracture. Pacemaker interrogation revealed a ventricular lead failing to sense or pace with high impedance. A pacemaker check 8 months previously was satisfactory.

Figure 1

Electrocardiogram on admission.

Figure 2

Left: chest radiograph showing distal lead fracture (arrow); right: acquisition at time of pacing showing distal lead fracture (arrow).

She underwent an uneventful new ventricular lead insertion and was discharged. The distal lead fracture (St Jude 1470T lead) was more evident during screening (figure 2, right). We were informed by St Jude Medical that this type of lead fracture is the first reported case documented with them worldwide. The patient died several months following discharge.

The incidence of pacemaker lead fracture has been reported to be as high as 7.2%,1 almost all of which occur in the infraclavicular region as a result of compression between the clavicle and first rib. To our knowledge this is the first reported case of this type of distal ventricular lead fracture. We can only postulate as to the mechanism of this incident. It was reported that over the preceding few months the patient had required increasing support while walking, and it may be this contributed to the lead fracture.